LI Zhen, WANG Hui, JIANG Yuwen. Laboratory Calibration of Phase-terrestrial 3D Laser Scanner[J]. Geomatics and Information Science of Wuhan University, 2014, 39(8): 987-991. DOI: 10.13203/j.whugis20130112
Citation: LI Zhen, WANG Hui, JIANG Yuwen. Laboratory Calibration of Phase-terrestrial 3D Laser Scanner[J]. Geomatics and Information Science of Wuhan University, 2014, 39(8): 987-991. DOI: 10.13203/j.whugis20130112

Laboratory Calibration of Phase-terrestrial 3D Laser Scanner

  • Objective Hardware-related system errors of terrestrial 3Dlaser scanner can be reduced or eliminatedby calibration.Basic model and reference model of terrestrial 3Dlaser scanner system errors were es-tablished.Laboratory calibration adjustment model was deduced.After the adjustment was comple-ted,exterior orientation elements,global coordinates of targets,coefficients of system error and ob-servations of corrected random error can be acquired.Considering both physically interpretable andempirically identified components,phase-terrestrial FARO Focus3Dself-calibration test was conduc-ted in a laboratory calibration room.The influence of the number and distribution of control points onthe residual distribution of observations and system error modeling were analyzed by calibration exper-iments.The modified model of FARO Focus3Dsystem errors was established.The effectiveness ofmodified model was proved.
  • loading

Catalog

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return